            Product name: Human HMGB-1(High Mobility Group Protein B1) ELISA Kit
            Reactivity: Human
            Alternative Names: HMGB1; HMG3; SBP1; Sulfoglucuronyl Carbohydrate Binding Protein; Amphoterin; High Mobility Group Box 1 Protein; HMG1; High Mobility Group Protein 1
            Assay Type: Sandwich
            Sensitivity: 26.9 pg/mL
            Standard: 4000 pg/mL
            Detection Range: 62.5-4000 pg/mL
            Sample Type: serum, plasma and other biological fluids
            Assay Length: 3.5h
            Research Area: Infection immunity;
            Uniprot ID: P09429
            Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human HMGB-1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human HMGB-1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human HMGB-1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human HMGB-1 in the samples is then determined by comparing the OD of the samples to the standard curve.

            標準曲線

            Concentration (pg/mL) OD Corrected OD
            4000.00 1.961 1.868
            2000.00 1.436 1.343
            1000.00 1.196 1.103
            500.00 0.841 0.748
            250.00 0.614 0.521
            125.00 0.415 0.322
            62.50 0.225 0.132
            0.00 0.093 0.000

            精密度

            Intra-assay Precision (Precision within an assay):CV%<8%

            Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

            Inter-assay Precision (Precision between assays):CV%<10%

            Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

            回收率

            Matrices listed below were spiked with certain level of recombinant HMGB-1 and the recovery rates were calculated by comparing the measured value to the expected amount of HMGB-1 in samples.
            Matrix Recovery range Average
            serum(n=5) 80-97% 88%
            EDTA plasma(n=5) 85-101% 93%
            Heparin plasma(n=5) 83-97% 90%

            線性

            The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HMGB-1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
            Matrix 1:2 1:4 1:8 1:16
            serum(n=5) 91-99% 83-89% 88-104% 97-103%
            EDTA plasma(n=5) 85-104% 83-96% 95-103% 88-95%
            Heparin plasma(n=5) 86-93% 93-101% 86-97% 90-101%